Dose reduction in 64-row whole-body CT in multiple trauma: an optimized CT protocol with iterative image reconstruction on a gemstone-based scintillator

多发性创伤中64排全身CT的剂量降低:基于宝石闪烁体的迭代图像重建优化CT方案

阅读:1

Abstract

OBJECTIVE: Evaluation of potential dose savings by implementing adaptive statistical iterative reconstruction (ASiR) on a gemstone-based scintillator in a clinical 64-row whole-body CT (WBCT) protocol after multiple trauma. METHODS: Dose reports of 152 WBCT scans were analysed for two 64-row multidetector CT scanners (Scanners A and B); the main scanning parameters were kept constant. ASiR and a gemstone-based scintillator were used in Scanner B, and the noise index was adjusted (head: 5.2 vs 6.0; thorax/abdomen: 29.0 vs 46.0). The scan length, CT dose index (CTDI) and dose-length product (DLP) were analysed. The estimated mean effective dose was calculated using normalized conversion factors. Student's t-test was used for statistics. RESULTS: Both the mean CTDI (mGy) (Scanner A: 53.8 ± 2.0, 10.3 ± 2.5, 14.4 ± 3.7; Scanner B: 48.7 ± 2.2, 7.1 ± 2.3, 9.1 ± 3.6; p < 0.001, respectively) and the mean DLP (mGy cm) (Scanner A: 1318.9 ± 167.8, 509.3 ± 134.7, 848.8 ± 254.0; Scanner B: 1190.6 ± 172.6, 354.6 ± 128.3, 561.0 ± 246.7; p < 0.001, respectively) for the head, thorax and abdomen were significantly reduced with Scanner B. There was no relevant difference in scan length. The total mean effective dose (mSv) was significantly decreased with Scanner B (24.4 ± 6.0, 17.2 ± 5.8; p < 0.001). CONCLUSION: The implementation of ASiR and a gemstone-based scintillator allows for significant dose savings in a clinical WBCT protocol. ADVANCES IN KNOWLEDGE: Recent technical developments can significantly reduce radiation dose of WBCT in multiple trauma. Dose reductions of 10-34% can be achieved.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。