Estimation of Surface Dose in the Presence of Unwanted Air Gaps under the Bolus in Postmastectomy Radiation Therapy: A Phantom Dosimetric Study

乳房切除术后放射治疗中,当剂量垫下方存在不必要的空气间隙时,表面剂量的估算:一项体模剂量学研究

阅读:1

Abstract

BACKGROUND: The aim of this study is to design and fabricate a thorax phantom with irregularly shaped trapezoidal slots across the left side of the chest wall, allowing for the creation of unwanted air gaps under the bolus. METHOD: Surface dose (Dsurf) measurements were made with Gaf Chromic EBT3 films at air gaps (0.0, 5.0, 10.0 and 15.0 mm) under gel bolus of thickness (5.0 mm & 10.0 mm), for 3DCRT technique (2 and 3 field) with clinical 6 MV photon beam under uniform and non-uniform air gap condition. The obtained values were compared with TPS estimated ones. RESULTS: In the presence of 15.0 mm uniform air gap, the mean estimated and measured Dsurf values with two and three field techniques decreased by 14.0 % to 15.2% and 14.7% to 17.4% under 5.0 mm and 10.0 mm bolus applications respectively. In presence of non-uniform air gap condition, the effect on Dsurf was minimal (3 to 3.5%) compared with the uniform air gap condition. CONCLUSIONS: Based on the study's findings, it is recommended that when using bolus in clinical radiotherapy applications, special care be taken to avoid unwanted air gaps under the bolus in order to achieve a uniform surface dose across the treatment region, where a customized 3D printed bolus may be a better option.