Dosimetric Validation of Volumetric Modulated Arc Therapy with Three 6MV Beam-Matched Linear Accelerators

利用三台6MV束匹配直线加速器对容积调强弧形治疗进行剂量学验证

阅读:1

Abstract

Background: To avoid inconvenience to patients due to linear accelerator down time in busy radio-therapy departments, treatment plans can be switched between linear accelerators provided that all exhibit the same same dosimetric characteristics. In other words linear accelerators should be beam-matched. Aim: The aim of this study was to evaluate the clinical significance of beam-matching using VMAT plans. Materials and Methods: Dosimetric data with a 6MV beam from am Clinac 2100CD were taken as baseline values and other two units, a 2300CD and a Unique Performance, were factory tuned in accordance. An analysis of PDD data was performed for different field sizes to evaluate energy matching. Beam profiles for field sizes of 10×10 cm2 and 40 × 40 cm2 at depths of 1.5 cm and 10 cm were analyzed. The relative output factor and MLC dosimetric properties were compared with each machine to determine variability among the different models. Thirty patients from our database were selected, ten each for head and neck, thorax and pelvis sites. VMAT plans were created in the Eclipse treatment planning system for a Clinac 2100 CD for reference. and verification plans were created for each to compare point dose measurements. Results: The TPR 20/10 for 10 × 10 cm2 was well matched, showing no energy differences. Deviation of all point dose measurements fell within ±3%. Planar dose maps all showed greater than 95% of points with a passed area γ-value less than 1. Conclusion: Our study evaluation of beam matching with treatment planning modeling showed good agreement fior 6 MV beams across all three linear accelerators used in our clinical environment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。