End-to-End QA with Polymer Gel Dosimeter for Photon Beam Radiation Therapy

采用聚合物凝胶剂量计进行光子束放射治疗的端到端质量保证

阅读:2

Abstract

With the complexity and high demands on quality assurance (QA) of photon beam radiation therapy, end-to-end (E2E) QA is necessary to validate the entire treatment workflow from pre-treatment imaging to beam delivery. A polymer gel dosimeter is a promising tool for three-dimensional (3D) dose distribution measurement. The purpose of this study is to design a fast "one delivery" polymethyl methacrylate (PMMA) phantom with a polymer gel dosimeter for the E2E QA test of the photon beam. The one delivery phantom is composed of ten calibration cuvettes for the calibration curve measurement, two 10 cm gel dosimeter inserts for the dose distribution measurement, and three 5.5 cm gel dosimeters for the square field measurement. The one delivery phantom holder is comparable in size and shape to that of a human thorax and abdomen. In addition, an anthropomorphic head phantom was employed to measure the patient-specific dose distribution of a VMAT plan. The E2E dosimetry was verified by undertaking the whole RT procedure (immobilization, CT simulation, treatment planning, phantom set-up, imaged-guided registration, and beam delivery). The calibration curve, field size, and patient-specific dose were measured with a polymer gel dosimeter. The positioning error can be mitigated with the one-delivery PMMA phantom holder. The delivered dose measured with a polymer gel dosimeter was compared with the planned dose. The gamma passing rate is 86.64% with the MAGAT-f gel dosimeter. The results ascertain the feasibility of the one delivery phantom with a polymer gel dosimeter for a photon beam in E2E QA. The QA time can be reduced with the designed one delivery phantom.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。