Effective DQE (eDQE) and speed of digital radiographic systems: an experimental methodology

数字放射成像系统的有效数字质量效率 (eDQE) 和速度:一种实验方法

阅读:1

Abstract

Prior studies on performance evaluation of digital radiographic systems have primarily focused on the assessment of the detector performance alone. However, the clinical performance of such systems is also substantially impacted by magnification, focal spot blur, the presence of scattered radiation, and the presence of an antiscatter grid. The purpose of this study is to evaluate an experimental methodology to assess the performance of a digital radiographic system, including those attributes, and to propose a new metric, effective detective quantum efficiency (eDQE), a candidate for defining the efficiency or speed of digital radiographic imaging systems. The study employed a geometric phantom simulating the attenuation and scatter properties of the adult human thorax and a representative indirect flat-panel-based clinical digital radiographic imaging system. The noise power spectrum (NPS) was derived from images of the phantom acquired at three exposure levels spanning the operating range of the clinical system. The modulation transfer function (MTF) was measured using an edge device positioned at the surface of the phantom, facing the x-ray source. Scatter measurements were made using a beam stop technique. The eDQE was then computed from these measurements, along with measures of phantom attenuation and x-ray flux. The MTF results showed notable impact from the focal spot blur, while the NPS depicted a large component of structured noise resulting from use of an antiscatter grid. The eDQE was found to be an order of magnitude lower than the conventional DQE. At 120 kVp, eDQE(0) was in the 8%-9% range, fivefold lower than DQE(0) at the same technique. The eDQE method yielded reproducible estimates of the system performance in a clinically relevant context by quantifying the inherent speed of the system, that is, the actual signal to noise ratio that would be measured under clinical operating conditions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。