A dosimetry study of post-mastectomy radiation therapy with AeroForm tissue expander

使用 AeroForm 组织扩张器进行乳房切除术后放射治疗的剂量学研究

阅读:1

Abstract

PURPOSE: To evaluate the dosimetric effects of the AeroForm(TM) (AirXanpders®, Palo Alto, CA) tissue expander in-situ for breast cancer patients receiving post-mastectomy radiation therapy. METHODS AND MATERIALS: A film phantom (P1) was constructed by placing the metallic canister of the AeroForm on a solid water phantom with EBT3 films at five depths ranging from 2.6 mm to 66.2 mm. A breast phantom (P2), a three-dimensional printed tissue-equivalent breast with fully expanded AeroForm in-situ, was placed on a thorax phantom. A total of 21 optical luminescent dosimeters (OLSDs) were placed on the anterior skin-gas interface and the posterior chest wall-metal interface of the AeroForm. Both phantoms were imaged with a 16-bit computed tomography scanner with orthopedic metal artifact reduction. P1 was irradiated with an open field utilizing 6 MV and 15 MV photon beams at 0°, 90°, and 270°. P2 was irradiated using a volumetric modulated arc therapy plan with a 6 MV photon beam and a tangential plan with a 15 MV photon beam. All doses were calculated using Eclipse (Varian, Palo Alto, CA) with AAA and AcurosXB (AXB) algorithms. RESULTS: The average dose differences between film measurements and AXB in the region adjacent to the canister in P1 were within 3.1% for 15 MV and 0.9% for 6 MV. Local dose differences over 10% were also observed. In the chest wall region of P2, the median dose of OLSDs in percentage of prescription dose were 108.4% (range 95.4%-113.0%) for the 15MV tangential plan and 110.4% (range 99.1%-113.8%) for the 6MV volumetric modulated arc therapy plan. In the skin-gas interface, the median dose of the OLSDs were 102.3% (range 92.7%-107.7%) for the 15 MV plan and 108.2% (range 97.8-113.5%) for the 6 MV plan. Measured doses were, in general, higher than calculated doses with AXB calculations. The AAA dose algorithms produced results with slightly larger discrepancies between measurements compared with AXB. CONCLUSIONS: The AeroForm creates significant dose uncertainties in the chest wall-metal interface. The AcurosXB dose calculation algorithm is recommended for more accurate calculations. If possible, post-mastectomy radiation therapy should be delivered after the permanent implant is in place.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。