Correlation between Carboxyhemoglobin Levels Measured by Blood Gas Analysis and by Multiwave Pulse Oximetry

血气分析法和多波脉搏血氧饱和度法测得的碳氧血红蛋白水平的相关性

阅读:1

Abstract

Carbon monoxide (CO) poisoning is difficult to diagnose owing to its nonspecific symptoms. Multiwave pulse oximetry can be used to quickly screen patients for CO poisoning. However, few studies have analyzed patients with CO poisoning who presented to the emergency department (ED). The primary aim of our study was to determine the correlation between COHb levels measured in blood gas analysis and COHb levels measured in multiwave pulse oximetry. Secondary aims were the sensitivity and specificity of the COHb level cutoff value using multiwave pulse oximetry to predict a 25% COHb level in blood gas analysis. This single-center retrospective observational study included patients with CO poisoning who visited the ED of a university-affiliated hospital in Seoul, Korea between July 2021 and June 2023. COHb poisoning was determined using blood gas analysis and multiwave pulse oximetry. The correlation of COHb levels between the two tests was evaluated using correlation analysis. The area under the receiver operating characteristic curve (AUC) of multiwave pulse oximetry was calculated to predict COHb levels from the blood gas analysis. The optimal cutoff values, sensitivity, and specificity of COHb were determined. A total of 224 patients who had COHb levels measured using both multiwave pulse oximetry and blood gas analysis were included in the analysis. In the correlation analysis, COHb showed a high positive correlation with COHb measured using blood gas analysis (Spearman correlation coefficient = 0.86, p < 0.001). The AUC of COHb measured by multiwave pulse oximetry to predict 25% of the COHb level (which can be an indication of hyperbaric oxygen treatment) measured by blood gas analysis was 0.916. When the COHb levels measured with multiwave pulse oximetry were 20% the sensitivity was 81% and the specificity was 83%, and when the COHb levels were 25% the sensitivity was 50% and the specificity was 95%. The COHb value measured using multiwave pulse oximetry blood gas analysis showed a high correlation. However, additional research using large-scale studies is required for validation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。