Influence of Curved Video Laryngoscope Blade Sizes and Patient Heights on Video Laryngoscopic Views: A Randomized Controlled Trial

弯曲型视频喉镜片尺寸和患者身高对视频喉镜视野的影响:一项随机对照试验

阅读:1

Abstract

This study aimed to compare the video laryngoscope views facilitated by curved blades 3 and 4 with an exploration of the relationship between these views and patient height. Conducted as a randomized controlled trial, this study enrolled adults scheduled for surgery under general anesthesia. Intubation procedures were recorded, and the percentage of glottic opening was measured before tube insertion. Multivariate analysis validated the impact of various factors, including blade size and patient height, on the percentage of glottic opening scores. A total of 192 patients were included. The median percentage of glottic opening scores for curved blades 3 and 4 were 100 and 83, respectively (p < 0.001). The unstandardized coefficient indicated a significant negative impact of blade 4 on the percentage of glottic opening scores (-13, p < 0.001). In the locally estimated scatterplot smoothing analysis, blade 3 exhibited a steady rise in glottic opening scores with increasing height, whereas blade 4 showed a peak followed by a decline around 185 cm. The unstandardized coefficient of height showed no significant association (0, p = 0.819). The study observed superior laryngoscopic views with blade 3 compared to blade 4. However, no significant association was found between laryngoscopic views and patient height.