PSI-Guided Mandible-First Orthognathic Surgery: Maxillo-Mandibular Position Accuracy and Vertical Dimension Adjustability

PSI引导下颌先行正颌手术:颌骨位置精度和垂直方向可调性

阅读:1

Abstract

In orthognathic surgery, patient-specific osteosynthesis implants (PSIs) represent a novel approach for the reproduction of the virtual surgical planning on the patient. The aim of this study is to analyse the quality of maxillo-mandibular positioning using a hybrid mandible-first mandibular-PSI-guided procedure on twenty-two patients while the upper maxilla was fixed using manually bent stock titanium miniplates. The virtual surgical plan was used to design PSIs and positioning guides, which were then 3D printed using biocompatible materials. A Cone Beam Computed Tomography (CBCT) scan was performed one month after surgery and postoperative facial skeletal models were segmented for comparison against the surgical plan. A three-dimensional cephalometric analysis was carried out on both planned and obtained anatomies. A Spearman correlation matrix was computed on the calculated discrepancies in order to achieve a more comprehensive description of maxillo-mandibular displacement. Intraoperatively, all PSIs were successfully applied. The procedure was found to be accurate in planned maxillo-mandibular positioning reproduction, while maintaining a degree of flexibility to allow for aesthetics-based verticality correction in a pitch range between -5.31 and +1.79 mm. Such a correction did not significantly affect the achievement of planned frontal symmetry.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。