Accuracy of Patient-Specific Drilling Guides in Acetabular Fracture Surgery: A Human Cadaver Study

髋臼骨折手术中患者特异性钻孔导板的准确性:一项人体尸体研究

阅读:1

Abstract

Due to the complex anatomical shape of the pelvis, screw placement can be challenging in acetabular fracture surgery. This study aims to assess the accuracy of screw placement using patient-specific surgical drilling guides applied to pre-contoured conventional implants in acetabular fracture surgery. CT scans were made of four human cadavers to create 3D models of each (unfractured) pelvis. Implants were pre-contoured on 3D printed pelvic models and optically scanned. Following virtual preoperative planning, surgical drilling guides were designed to fit on top of the implant and were 3D printed. The differences between the pre-planned and actual screw directions (degrees) and screw entry points (mm) were assessed from the pre- and postoperative CT-scans. The median difference between the planned and actual screw direction was 5.9° (IQR: 4-8°) for the in-plate screws and 7.6° (IQR: 6-10°) for the infra-acetabular and column screws. The median entry point differences were 3.6 (IQR: 2-5) mm for the in-plate screws and 2.6 (IQR: 2-3) mm for the infra-acetabular and column screws. No screws penetrated into the hip joint or caused soft tissue injuries. Three-dimensional preoperative planning in combination with surgical guides that envelope pre-contoured conventional implants result in accurate screw placement during acetabular fracture surgery.