Free Myocutaneous Flap Assessment in a Rat Model: Verification of a Wireless Bioelectrical Impedance Assessment (BIA) System for Vascular Compromise Following Microsurgery

在大鼠模型中评估游离肌皮瓣:验证无线生物电阻抗评估(BIA)系统对显微手术后血管损伤的评估

阅读:1

Abstract

BACKGROUND: Microvascular tissue transfer is a common reconstructive procedure. We designed a bioelectrical impedance assessment (BIA) system for quantitative analysis of tissue status. This study attempts to verify it through the animal model. METHODS: The flaps of the rat model were monitored by the BIA system. RESULTS: The BIA variation of the free flap in the rat after the vascular compromise was recorded. The non-vascular ligation limbs of the same rat served as a control group. The bio-impedance in the experimental group was larger than the control group. The bio-impedances of both the thigh/feet flaps in the experimental group were increased over time. In the thigh, the difference in bio-impedance from the control group was first detected at 10 kHz at the 3rd and last at 1 kHz at the 6th h, after vascular compromise. The same finding was observed in the feet. Compared with the control group, the bio-impedance ratio (1 kHz/20 kHz) of the experimental group decreased with time, while their variation tendencies in the thigh and feet were similar. CONCLUSIONS: The flap may be monitored by the BIA for vascular status.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。