Biomechanical analysis of the door-shaped titanium plate in single-level anterior cervical discectomy and fusion

单节段前路颈椎间盘切除融合术中门形钛板的生物力学分析

阅读:1

Abstract

BACKGROUND: Analyse and discuss the immediate stability of the cervical spine after anterior cervical discectomy and fusion using a door-shaped titanium plate and compare it with the traditional titanium plate, to provide biomechanical evidence for the rationality and effectiveness of the door-shaped titanium plate in clinical applications. METHODS: Ten adult goat C4/5 vertebral bodies were obtained, and models were prepared using denture base resin. Biomechanical experiments were performed on the specimens before internal fixation. MTS was used to conduct non-destructive biomechanical loading tests in six directions, including flexion, extension, left-right bending, and left-right torsion, recording the range of motion (ROM) and neutral zone (NZ) of each specimen. The specimens were then randomly divided into two groups: the study group was fixed with a door-shaped titanium plate, and the control group was fixed with a traditional titanium plate. ROM and NZ in each direction were measured again. After measurements, both groups were subjected to 0.5 Hz torsion loading with a torque of 2 N m for a total of 3000 cycles, followed by measuring ROM and NZ in six directions once more. RESULTS: Compared to before fixation, ROM and NZ in both groups significantly decreased in all six directions after fixation, with statistical significance (P < 0.05); after fixation, the study group showed slightly lower values for various mechanical reference parameters compared to the control group, with no statistical significance (P > 0.05); after 3000 torsional loads, both internal fixation groups showed increased ROM and NZ compared to after fixation but to a lower extent, and no screw or titanium plate loosening was observed. Compared to before fixation, the differences were still statistically significant (P < 0.05), with the study group having slightly lower ROM and NZ values in all directions compared to the control group, with no statistical significance (P > 0.05). CONCLUSION: The door-shaped titanium plate exhibits mechanical properties similar to the traditional titanium plate in all directions, and its smaller size and simpler surgical operation can be used for anterior cervical endoscopic surgery, reducing surgical trauma. It is clinically feasible and deserves further research and promotion.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。