An accelerometer-based navigation system provides acetabular cup orientation accuracy comparable to that of computed tomography-based navigation during total hip arthroplasty in the supine position

在仰卧位全髋关节置换术中,基于加速度计的导航系统可提供与基于计算机断层扫描的导航系统相当的髋臼杯定位精度。

阅读:1

Abstract

BACKGROUND: Inadequate acetabular component orientation is associated with postoperative impingement, dislocation, and accelerated polyethylene wear. Computed tomography (CT)-based navigation systems provide accuracy for total hip arthroplasty (THA) but are not available in all facilities. Accelerometer-based navigation systems are inexpensive, but their accuracy remains undetermined. This study compares the accuracy of cup orientation in THA using CT-based and accelerometer-based navigation systems. METHODS: This retrospective study included 35 consecutive patients (11 males, 24 females; mean age, 65 years) who underwent primary cementless THA via an anterolateral approach in the supine position. Both CT-based and accelerometer-based navigation systems were used simultaneously. The accuracy of cup orientation was compared between the two systems using postoperative CT. RESULTS: The accuracy of cup inclination was 2.7° ± 2.0° in the CT-based group and 3.3° ± 2.4° in the accelerometer-based group. The accuracy of cup anteversion was 2.8° ± 2.6° in the CT-based group and 3.4° ± 2.2° in the accelerometer-based group. No significant difference was observed in cup inclination (p = 0.29) or cup anteversion (p = 0.34) between CT-based and accelerometer-based navigation. CONCLUSIONS: The accuracy of cup positioning did not differ significantly between CT-based and accelerometer-based navigation systems.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。