Numerical Study of Fabrication-Related Effects of the Structural-Profile on the Performance of a Dielectric Photonic Crystal-Based Fluid Sensor

介电光子晶体流体传感器结构轮廓对器件性能的加工相关影响的数值研究

阅读:1

Abstract

In this work, fabrication of a dielectric photonic crystal device and numerical study of its spectral characteristics as a refractive index sensor are presented for near infrared range. The proposed nanosensor device is composed of low-cost dielectric materials, i.e., silicon dioxide and niobium pentoxide, and is fabricated using focused ion-beam milling lithography. In the first part, the fabrication process of the device is discussed, along with the process parameters and their effects on the structural properties of the resulting photonic crystal elements. In the second part, the device is numerically tested as a sensor for the biological refractive index range of 1.33 to 1.4. The performance considerations of the biosensor device are studied for 12 different structural profiles based on the fabrication results. It is shown that the angular-wall-profile of the fabricated structures downgrades the performance of the sensor, and the optimum value of hole depth should be in the range of 930-1500 nm to get the best performance. A sensitivity of 185.117 nm/RIU and a figure of merit of 9.7 were recorded for the optimum design of the device; however, a maximum sensitivity of 296.183 nm/RIU and a figure-of-merit of 13.184 RIU(-1) were achieved. The device is recommended for a variety of biosensing applications due to its inert material properties, stable design and easy integration with fiber-optic setups.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。