Simultaneous Multiplexed Detection of Protein and Metal Ions by a Colorimetric Microfluidic Paper-based Analytical Device

基于比色微流控纸基分析装置的蛋白质和金属离子同时多重检测

阅读:1

Abstract

In order to improve the efficiency of disease diagnosis and environmental monitoring, it is desirable to detect the concentration of proteins and metal ions simultaneously, since the current popular diagnostic platform can only detect proteins or metal ions independently. In this work, we developed a colorimetric microfluidic paper-based analytical device (µPAD) for simultaneous determination of protein (bovine serum albumin, BSA) and metal ions [Fe(III) and Ni(II)]. The µPAD consisted of one central zone, ten reaction zones and ten detection zones in one device, in which reaction solutions were effectively optimized for different types of chromogenic reactions. Fe(III), Ni(II) and BSA can be easily identified by the colored products, and their concentrations are in good accordance with color depth based on the established standard curves. The detection limits are 0.1 mM for Fe(III), 0.5 mM for Ni(II) and 1µM for BSA, respectively. Best of all, we demonstrated the efficiency of the µPAD with accurate detection of Fe(III), Ni (II) and BSA from river water samples within 15 minutes. The µPAD detection is efficient, instrument-free, and easy-to-use, holding great potential for simultaneous detection of cross type analytes in numerous diagnostic fields.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。