Detection of Paroxysms in Long-Term, Single-Channel EEG-Monitoring of Patients with Typical Absence Seizures

长期单通道脑电图监测典型失神发作患者阵发性发作的检测

阅读:1

Abstract

Absence seizures are associated with generalized 2.5-5 Hz spike-wave discharges in the electroencephalogram (EEG). Rarely are patients, parents, or physicians aware of the duration or incidence of seizures. Six patients were monitored with a portable EEG-device over four times 24 h to evaluate how easily outpatients are monitored and how well an automatic seizure detection algorithm can identify the absences. Based on patient-specific modeling, we achieved a sensitivity of 98.4% with only 0.23 false detections per hour. This yields a clinically satisfying performance with a positive predictive value of 87.1%. Portable EEG-recorders identifying paroxystic events in epilepsy outpatients are a promising tool for patients and physicians dealing with absence epilepsy. Albeit the small size of the EEG-device, some children still complained about the obtrusive nature of the device. We aim at developing less obtrusive though still very efficient devices, e.g., hidden in the ear canal or below the skin.