Development and Validation of a Quantitative LC-MS/MS Method for CDK4/6 Inhibitors Palbociclib, Ribociclib, Abemaciclib, and Abemaciclib-M2 in Human Plasma

开发并验证人血浆中 CDK4/6 抑制剂 Palbociclib、Ribociclib、Abemaciclib 和 Abemaciclib-M2 的定量 LC-MS/MS 方法

阅读:7
作者:Sarah M Burke, Mustafa Kamal, Andrew K L Goey

Background

The cyclin-dependent kinase 4 and 6 (CDK4/6) inhibitors, palbociclib, ribociclib, and abemaciclib, are standard-of-care agents for patients with hormone receptor-positive human epidermal growth factor receptor 2-negative metastatic breast cancer. In support of therapeutic drug monitoring and clinical pharmacokinetic studies, a liquid chromatography coupled with tandem mass spectrometry assay for the simultaneous quantitation of CDK4/6 inhibitors and the major active metabolite M2 of abemaciclib in human plasma has been developed.

Conclusions

A sensitive liquid chromatography coupled with tandem mass spectrometry method to measure CDK4/6 inhibitors was developed and validated according to the Food and Drug Administration criteria. Quantitation of all analytes in clinical plasma samples confirmed that the assay is suitable for therapeutic drug monitoring and clinical pharmacokinetic studies of CDK4/6 inhibitors.

Methods

Analytes were extracted from 50 μL of human plasma by precipitating proteins with methanol and then collecting the supernatant. Reversed-phase high-performance liquid chromatography was performed for analyte separation using a biphasic gradient at a flow rate of 0.25-0.5 mL/min. The total run time was 9.5 minutes. The analytes were detected using MS/MS with electrospray ionization operating in positive ion mode.

Results

Validation according to the US Food and Drug Administration's guidance showed that the new assay produced accurate (94.7%-107%) and precise (within-run: 1.2%-8.2%; between-run: 0.6%-7.5%) measurements of all analytes over a concentration range of 5-2000 ng/mL. Overall, analyte recoveries were consistent (mean values: 110%-129%). The analytes were also stable in human plasma and the final extract under various storage conditions. Finally, the clinical applicability of the assay was confirmed by quantitation of all analytes in plasma samples obtained from patients treated with CDK4/6 inhibitors. Reproducibility of the measured analyte concentrations in study samples was confirmed successfully by incurred sample reanalysis. Conclusions: A sensitive liquid chromatography coupled with tandem mass spectrometry method to measure CDK4/6 inhibitors was developed and validated according to the Food and Drug Administration criteria. Quantitation of all analytes in clinical plasma samples confirmed that the assay is suitable for therapeutic drug monitoring and clinical pharmacokinetic studies of CDK4/6 inhibitors.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。