Multicenter Evaluation of the Vitek MS v3.0 System for the Identification of Filamentous Fungi

Vitek MS v3.0系统在丝状真菌鉴定中的多中心评价

阅读:1

Abstract

Invasive fungal infections are an important cause of morbidity and mortality affecting primarily immunocompromised patients. While fungal identification to the species level is critical to providing appropriate therapy, it can be slow and laborious and often relies on subjective morphological criteria. The use of matrix-assisted laser desorption ionization-time of flight (MALDI-TOF) mass spectrometry has the potential to speed up and improve the accuracy of identification. In this multicenter study, we evaluated the accuracy of the Vitek MS v3.0 system in identifying 1,601 clinical mold isolates compared to identification by DNA sequence analysis and supported by morphological and phenotypic testing. Among the 1,519 isolates representing organisms in the v3.0 database, 91% (n = 1,387) were correctly identified to the species level. An additional 27 isolates (2%) were correctly identified to the genus level. Fifteen isolates were incorrectly identified, due to either a single incorrect identification (n = 13) or multiple identifications from different genera (n = 2). In those cases, when a single identification was provided that was not correct, the misidentification was within the same genus. The Vitek MS v3.0 was unable to identify 91 (6%) isolates, despite repeat testing. These isolates were distributed among all the genera. When considering all isolates tested, even those that were not represented in the database, the Vitek MS v3.0 provided a single correct identification 98% of the time. These findings demonstrate that the Vitek MS v3.0 system is highly accurate for the identification of common molds encountered in the clinical mycology laboratory.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。