Residual decontamination chemical agents negatively affect adhesion and proliferation of osteoblast-like cells on implant surface

残留的净化化学药剂对植入物表面成骨细胞样细胞的粘附和增殖产生负面影响

阅读:10
作者:Ísis de Fátima Balderrama, Matheus Völz Cardoso, Vitor Toledo Stuani, Rodrigo Cardoso Oliveira, Adriana Arruda Matos, Sebastião Luiz Aguiar Greghi, Adriana Campos Passanezi Sant'Ana

Conclusions

These findings suggest that decontaminated surfaces were able to impair the counting of osteoblast-like cell adhesion and proliferation.

Methods

Commercially available implants of different brands and surface characteristics were selected: Biomet 3i® Nanotite (NT) and Osseotite (OT), Straumann® SLActive (SLA), and Neodent® Acqua Drive (ACQ) and Neoporos Drive CM (CM). Physical and chemical properties of the implants were investigated by scanning electron microscopy (SEM), energy dispersive X-ray spectroscopy (EDS), and wettability analysis (WETT). Implants were previously contaminated with Aggregatibacter actinomycetemcomitans strains; after that, samples were decontaminated by different chemical methods. Decontaminated (test group; n = 15/type of implant) and uncontaminated (control group; n = 5/type of implant) samples were analyzed according to the number of human osteoblastic osteosarcoma cells (Saos-2) adhered on the implant surface after 24 h and 72 h in SEM images.

Purpose

To investigate the influence of implant surface decontaminated and uncontaminated on osteoblast-like cell adhesion and proliferation MATERIALS AND

Results

ACQ was found to be highly hydrophilic, and NT was the most hydrophobic implant. Increased variation of Saos-2 cell adhesion and proliferation were observed on all test and control groups. Controversially, at the proliferation analysis in 72 h, CM implant was the only implant that showed no significant difference between test and group (p = 0.2833; Tukey's multiple comparisons test). NT implants showed the greater value of cell proliferation when compared with all types of implant surface (p = 0.0002; Tukey's multiple comparisons test). Conclusions: These findings suggest that decontaminated surfaces were able to impair the counting of osteoblast-like cell adhesion and proliferation.