A simple and accurate LC‑MS/MS method for monitoring cyclosporin A that is suitable for high throughput analysis

一种简便、准确的LC-MS/MS方法,适用于高通量分析,可用于监测环孢素A。

阅读:2

Abstract

With time, the number of samples in clinical laboratories from therapeutic drug monitoring has increased. Existing analytical methods for blood cyclosporin A (CSA) monitoring, such as high-performance liquid chromatography (HPLC) and immunoassays, have limitations including cross-reactivity, time consumption, and the complicated procedures involved. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) has long been considered the reference standard owing to its high accuracy, specificity, and sensitivity. However, large numbers of blood samples, multi-step preparation procedures, and longer analytical times (2.5-20 min) are required as a consequence of the different technical strategies, to ensure good analytical performance and routine quality assurance. A stable, reliable, and high throughput detection method will save personnel time and reduce laboratory costs. Therefore, a high throughput and simple LC-MS/MS method was developed and validated for the detection of whole-blood CSA with CSA-d12 as the internal standard in the present study. Whole blood samples were prepared through a modified one-step protein precipitation method. A C18 column (50x2.1 mm, 2.7 µm) with a mobile phase flow rate of 0.5 ml/min was used for chromatographic separation with a total running time of 4.3 min to avoid the matrix effect. To protect the mass spectrometer, only part of the sample after LC separation was allowed to enter the mass spectrum, using two HPLC systems coupled to one mass spectrometry. In this way, throughput was improved with detection of two samples possible within 4.3 min using a shorter analytical time for each sample of 2.15 min. This modified LC-MS/MS method showed excellent analytical performance and demonstrated less matrix effect and a wide linear range. The design of multi-LC systems coupled with one mass spectrometry may play a notable role in the improvement of daily detection throughput, speeding up LC-MS/MS, and allowing it to be an integral part of continuous diagnostics in the near future.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。