Implantable collamer lens with a central hole for residual refractive error correction after corneal refractive surgery

用于矫正角膜屈光手术后残余屈光不正的带中心孔的可植入式胶原蛋白镜片

阅读:1

Abstract

The present study aimed to assess the visual and refractive outcomes of an implantable collamer lens with a central hole (ICL V4c) for residual refractive error correction after corneal refractive surgery in individuals with myopia. A total of 16 eyes of eight consecutive patients with myopia undergoing ICL V4c implantation after corneal refractive surgery were investigated. The uncorrected visual acuity (VA) and best-corrected VA were examined prior to surgery and at 1, 3 and 6 months after surgery. The post-operative values of the modulation transfer function (MTF) cutoff frequency, Strehl ratio (SR), objective scattering index (OSI) and Optical Quality Analysis System (OQAS) values (OVs) were quantitatively assessed using an OQAS. At 6 months after surgery, the mean uncorrected LogMAR VA was 0.06±0.10 and the values had improved in 100% of the eyes. The mean MTF cutoff frequency, SR, OSI, OV 100%, OV 20% and OV 9%, were 31.294±4.321 cycles/degree, 0.187±0.039, 1.399±0.274, 1.066±0.261, 0.748±0.287 and 0.509±0.229, respectively. In conclusion, ICL V4c implantation for the correction of residual refractive error after corneal refractive surgery appeared feasible and safe and also had an excellent optical performance. However, long-term changes in visual quality require further investigation.