Selective Extraction of a Monoisotopic Ion While Keeping the Other Ions in Flight on a Multi-Turn Time-of-Flight Mass Spectrometer

在多圈飞行时间质谱仪上选择性提取单同位素离子,同时保持其他离子在飞行中

阅读:1

Abstract

Using a multi-turn time-of-flight (TOF) mass spectrometer, we have extracted a single xenon isotope ion, (129)Xe(+), from its orbit at given a lap number without disturbing the rest of isotopes. After detecting the (129)Xe(+) at 20 laps, the rest of the xenon isotope spectrum was obtained at 30 laps, which generated a TOF spectrum where the TOF difference between (129)Xe(+) and (130)Xe(+) was 87.4 μs while (130)Xe(+) and (131)Xe(+) were 1.03 μs. The time distance between (129)Xe(+) and other isotopes can be set by any lap difference that is a factor of 8.7 μs, which depends on the acceleration voltage and the mass of the ion. Method accuracy was verified by comparing the isotopic abundance ratio of the xenon sample after withdrawing one of the ions from the isotope cluster to the abundance ratio obtained from the conventional method. The TOF stability was also evaluated at various lap numbers between 10 to 230.