Effect of ultrasound transducer face reflectivity on the light fluence inside a turbid medium in photoacoustic imaging

超声换能器表面反射率对光声成像中浑浊介质内部光通量的影响

阅读:1

Abstract

Many endoscopic, intravascular, and transvaginal applications require light to be delivered through optical fibers in a reflection mode. For photoacoustic imaging in reflection geometry, the front-face reflectivity of the ultrasound transducer face imposes a boundary condition that affects the light fluence and its distribution inside a turbid medium. Understanding and characterizing this boundary condition is critical for maximizing tissue illumination and therefore the signal-to-noise ratio of the photoacoustic signal. We systematically analyze the light fluence under three typical commercial transducer faces having reflection coefficients of 1.4, 18, and 28%, and compare the results to a transducer face with 60% coefficient at the laser wavelength of 750 nm. Monte Carlo simulations and experimental results show that light fluence and distribution obtained inside a turbid medium with the use of the 60% reflection coefficient transducer face has a significant improvement over the others, especially at shallower depths.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。