Abstract

Objective: The aim of this study is to evaluate accuracy of research activity monitors in measuring steps in older adults with a range of walking abilities. Method: Participants completed an initial assessment of gait speed. The accuracy of each monitor to record 100 steps was assessed across two walking trials. Results: In all, 43 older adults (age 87 ± 5.7 years, 81.4% female) participated. Overall, the StepWatch had the highest accuracy (99.0% ± 1.5%), followed by the ActivPAL (93.7% ± 11.1%) and the Actigraph (51.4% ± 35.7%). The accuracy of the Actigraph and ActivPAL varied according to assistive device use, and the accuracy of all three monitors differed by gait speed category (all p < .05). StepWatch was highly accurate (⩾97.7) across all conditions. Discussion: The StepWatch and ActivPAL monitor were reasonably accurate in measuring steps in older adults who walk slowly and use an assistive device. The Actigraph significantly undercounted steps in those who walk slow or use an assistive device. Researchers should consider gait speed and the use of assistive devices when selecting an activity monitor.