Ultralow Limit Detection of Soluble HER2 Biomarker in Serum with a Fiber-Optic Ball-Tip Resonator Assisted by a Tilted FBG

利用倾斜光纤布拉格光栅辅助的光纤球尖谐振器实现血清中可溶性HER2生物标志物的超低检测限

阅读:1

Abstract

An optical-fiber biosensor has been developed for the detection of the breast cancer biomarker soluble human epidermal growth factor receptor-2 (sHER2). The sensor was fabricated by combining a tilted fiber Bragg grating (TFBG) with a ball resonator, allowing us to achieve an excellent sensitivity compared to other optical-fiber-based sensors. The sensor exhibits a resonance comb excited by the TFBG and the spectral profile of the ball resonator. The detection of sHER2 at extremely low concentrations was carried out by tracking the amplitude change of selected resonances. The therapeutic anti-HER2 monoclonal antibody Trastuzumab has been used to functionalize the biosensor with silane surface chemistry. The sensor features a sensitivity of 4034 dB/RIU with a limit of detection (LoD) in buffer and in a 1/10 diluted serum of 151.5 ag/mL and 3.7 pg/mL, respectively. At relatively high protein concentrations (64 ng/mL) binding to sHER (7.36 dB) as compared to control proteins (below 0.7 dB) attested the high specificity of sHER2 detection.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。