Evaluation of anaerobic microdilution MIC results with the Prompt Inoculation System

利用快速接种系统评估厌氧微量稀释法MIC结果

阅读:1

Abstract

The Prompt Inoculation System (3M Co.) was compared with the overnight suspension inoculation procedure used with a broth microdilution anaerobic commercial system (Micro-Media Systems) for differences in MIC. MIC results from both suspension methods using six National Committee for Clinical Laboratory Standards-recommended quality control organisms were identical in 18 instances (75%) and within +/- 1 log2 dilution in 96% of the comparisons. Results with 45 anaerobic clinical isolates also compared satisfactorily; 83% of the results were identical and 97% were within +/- 1 log2 dilution. In addition, the direct (Prompt)-inoculated microdilution trays produced better growth and more valid MIC results; 92% of the clinical isolates produced MIC results versus 79% by the overnight suspension procedure.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。