Sub-parts-per-trillion level sensitivity in trace gas detection by cantilever-enhanced photo-acoustic spectroscopy

悬臂增强光声光谱法检测痕量气体的亚万亿级灵敏度

阅读:4
作者:Teemu Tomberg, Markku Vainio, Tuomas Hieta, Lauri Halonen

Abstract

An exceptional property of photo-acoustic spectroscopy is the zero-background in wavelength modulation configuration while the signal varies linearly as a function of absorbed laser power. Here, we make use of this property by combining a highly sensitive cantilever-enhanced photo-acoustic detector, a particularly stable high-power narrow-linewidth mid-infrared continuous-wave optical parametric oscillator, and a strong absorption cross-section of hydrogen fluoride to demonstrate the ability of cantilever-enhanced photo-acoustic spectroscopy to reach sub-parts-per-trillion level sensitivity in trace gas detection. The high stability of the experimental setup allows long averaging times. A noise equivalent concentration of 650 parts-per-quadrillion is reached in 32 minutes.