(1) H-(13) C independently tuned radiofrequency surface coil applied for in vivo hyperpolarized MRI

(1)H-(13)C独立调谐射频表面线圈应用于体内超极化磁共振成像

阅读:1

Abstract

PURPOSE: To develop a lump-element double-tuned common-mode-differential-mode (CMDM) radiofrequency (RF) surface coil with independent frequency tuning capacity for MRS and MRI applications. METHODS: The presented design has two modes that can operate with different current paths, allowing independent frequency adjustment. The coil prototype was tested on the bench and then examined in phantom and in vivo experiments. RESULTS: Standard deviations of frequency and impedance fluctuations measured in one resonator, while changing the tuning capacitor of another resonator, were less than 13 kHz and 0.55 Ω. The unloaded S21 was -36 dB and -41 dB, while the unloaded Q factor was 260 and 287, for (13) C and (1) H, respectively. In vivo hyperpolarized (13) C MR spectroscopy data demonstrated the feasibility of using the CMDM coil to measure the dynamics of lactate, alanine, pyruvate and bicarbonate signal in a normal rat head along with acquiring (1) H anatomical reference images. CONCLUSION: Independent frequency tuning capacity was demonstrated in the presented lump-element double-tuned CMDM coil. This CMDM coil maintained intrinsically decoupled magnetic fields, which provided sufficient isolation between the two resonators. The results from in vivo experiments demonstrated high sensitivity of both the (1) H and (13) C resonators. Magn Reson Med 76:1612-1620, 2016. © 2015 International Society for Magnetic Resonance in Medicine.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。