Prospective motion correction using coil-mounted cameras: Cross-calibration considerations

利用线圈式摄像机进行前瞻性运动校正:交叉校准考虑因素

阅读:1

Abstract

PURPOSE: Optical prospective motion correction substantially reduces sensitivity to motion in neuroimaging of human subjects. However, a major barrier to clinical deployment has been the time-consuming cross-calibration between the camera and MRI scanner reference frames. This work addresses this challenge. METHODS: A single camera was mounted onto the head coil for tracking head motion. Two new methods were developed: (1) a rapid calibration method for camera-to-scanner cross-calibration using a custom-made tool incorporating wireless active markers, and (2) a calibration adjustment method to compensate for table motion between scans. Both methods were tested at 1.5T and 3T in vivo. Simulations were performed to determine the required mechanical tolerance for repositioning of the camera. RESULTS: The rapid calibration method is completed in a short (<30 s) scan, which is carried out only once per installation. The calibration adjustment method requires no extra scan time and runs automatically whenever the system is used. The mechanical tolerance analysis indicates that most motion (90% reduction in voxel displacement) could be corrected even with far larger camera repositioning errors than are observed in practice. CONCLUSION: The methods presented here allow calibration of sufficient quality to be carried out and maintained with no additional technologist workload. Magn Reson Med 79:1911-1921, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。