Performance of a fast and high-resolution multi-echo spin-echo sequence for prostate T(2) mapping across multiple systems

多种系统下快速高分辨率多回波自旋回波序列在前列腺T2映射中的性能

阅读:1

Abstract

PURPOSE: To evaluate the performance of a multi-echo spin-echo sequence with k-t undersampling scheme (k-t T(2) ) in prostate cancer. METHODS: Phantom experiments were performed at five systems to estimate the bias, short-term repeatability, and reproducibility across all systems expressed with the within-subject coefficient of variation (wCV). Monthly measurements were performed on two systems for long-term repeatability estimation. To evaluate clinical repeatability, two T(2) maps (voxel size 0.8 × 0.8 × 3 mm(3) ; 5 min) were acquired at separate visits on one system for 13 prostate cancer patients. Repeatability was assessed per patient in relation to spatial resolution. T(2) values were compared for tumor, peripheral zone, and transition zone. RESULTS: Phantom measurements showed a small bias (median = -0.9 ms) and good short-term repeatability (median wCV = 0.5%). Long-term repeatability was 0.9 and 1.1% and reproducibility between systems was 1.7%. The median bias observed in patients was -1.1 ms. At voxel level, the median wCV was 15%, dropping to 4% for structures of 0.5 cm(3) . The median tumor T(2) values (79 ms) were significantly lower (P < 0.001) than in the peripheral zone (149 ms), but overlapped with the transition zone (91 ms). CONCLUSIONS: Reproducible T(2) mapping of the prostate is feasible with good spatial resolution in a clinically reasonable scan time, allowing reliable measurement of T(2) in structures as small as 0.5 cm(3) . Magn Reson Med 79:1586-1594, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。