Ultrafast compartmentalized relaxation time mapping with linear algebraic modeling

利用线性代数模型进行超快区室化弛豫时间映射

阅读:1

Abstract

PURPOSE: To dramatically accelerate compartmental-average longitudinal (T(1) ) and transverse (T(2) ) relaxation measurements using the minimal-acquisition linear algebraic modeling (SLAM) method, and to validate it in phantoms and humans. METHODS: Relaxation times were imaged at 3 Tesla in phantoms, in the abdomens of six volunteers, and in six brain tumor patients using standard inversion recovery and multi-spin-echo sequences. k-space was fully sampled to provide reference T(1) and T(2) measurements, and SLAM was performed using a limited set of phase encodes from central k-space. Anatomical compartments were segmented on scout images post-acquisition, and SLAM reconstruction was implemented using two algorithms. Compartment-average T(1) and T(2) measurements were determined retroactively from fully sampled data sets, and proactively from SLAM data sets at acceleration factors of up to 16. Values were compared with reference measurements. The compartment's localization properties were analyzed using the discrete spatial response function. RESULTS: At 16-fold acceleration, compartment-average SLAM T(1) measurements agreed with the full k-space compartment-average results to within 0.0% ± 0.7%, 1.4% ± 3.4%, and 0.5% ± 2.9% for phantom, abdominal, and brain T(1) measurements, respectively. The corresponding T(2) measurements agreed within 0.2% ± 1.9%, 0.9% ± 7.9%, and 0.4% ± 5.8%, respectively. CONCLUSION: SLAM can dramatically accelerate relaxation time measurements when compartmental or lesion-average values can suffice, or when standard relaxometry is precluded by scan-time limitations. Magn Reson Med 79:286-297, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。