Syncope triggered by atrial flutter in a patient with a pacemaker: cross-stimulation-a case report

起搏器植入患者因心房扑动诱发晕厥:交叉刺激——病例报告

阅读:1

Abstract

BACKGROUND: Syncope in a patient with a pacemaker is a serious event requiring urgent action to ascertain its cause. Around 5% of cases are due to a pacemaker system malfunction. CASE SUMMARY: An 82-year-old man underwent dual-chamber permanent pacemaker implantation due to intermittent high-degree atrioventricular block (AVB) in sinus rhythm. Nine months later, the patient reported episodes of syncope. The chest X-ray showed both leads to be at their expected positions. The electrocardiography (ECG) showed common atrial flutter. Ventricular capture during pacing in atrial demand pacing (AAI) mode confirmed cross-stimulation due to the switching of the atrial and ventricular leads at the pacemaker header. DISCUSSION: Cross-stimulation is a rare possibility in a differential diagnosis of causes of syncope. The diagnosis is frequently made during the procedure or a few hours later. The lack of symptoms during 9 months in this case was likely due to the patient having normal sinus rhythm with preserved AV conduction most of the time, as well as ventricular capture from the atrial lead related to non-sensed P waves. When atrial arrhythmias occurred, the sensing of the F waves inhibited ventricular pacing. In order to avoid this complication, in patients with intermittent bradycardia, pacing at a slightly higher heart rate during implantation of the device should be recommended to see the chamber paced with the surface ECG connected to the device interrogator. The ECG and electrogram (EGM) should correlate during device interrogation in order to identify this complication.).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。