Objective assessment of task performance: a comparison of two FFDM detectors using an anthropomorphic breast phantom

任务表现的客观评估:使用拟人乳房模型对两个 FFDM 探测器进行比较

阅读:8
作者:Andrey Makeev, Lynda C Ikejimba, Jesse Salad, Stephen J Glick

Abstract

Current digital mammography systems primarily employ one of two types of detectors: indirect conversion, typically using a cesium-iodine scintillator integrated with an amorphous silicon photodiode matrix, or direct conversion, using a photoconductive layer of amorphous selenium (a-Se) combined with thin-film transistor array. The goal of this study was to evaluate a methodology for objectively assessing image quality to compare human observer task performance in detecting microcalcification clusters and extended mass-like lesions achieved with different detector types. The proposed assessment methodology uses a novel anthropomorphic breast phantom fabricated with ink-jet printing. In addition to human observer detection performance, standard linear metrics such as modulation transfer function, noise power spectrum, and detective quantum efficiency (DQE) were also measured to assess image quality. An Analogic Anrad AXS-2430 a-Se detector used in a commercial FFDM/DBT system and a Teledyne Dalsa Xineos-2329 with CMOS pixel readout were evaluated and compared. The DQE of each detector was similar over a range of exposures. Similar task performance in detecting microcalcifications and masses was observed between the two detectors over a range of clinically applicable dose levels, with some perplexing differences in the detection of microcalcifications at the lowest dose measurement. The evaluation approach presented seems promising as a new technique for objective assessment of breast imaging technology.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。