The Cleaner XT™ Device as an Endovascular Adjunct for Pharmacomechanical Thrombolysis of Thrombosed Arteriovenous Fistulas and Grafts

Cleaner XT™装置作为血管内辅助手段用于血栓形成动静脉瘘和移植血管的药物机械溶栓治疗

阅读:1

Abstract

Objective: This paper documents our experience using the Cleaner XT™ device (Argon Medical Devices, Plano, TX, USA) for pharmacomechanical thrombolysis (PMT) of thrombosed haemodialysis arteriovenous grafts (AVG) and fistulas (AVF). Materials and Methods: This was a retrospective case series (n=17) over six months at Singapore General Hospital. We evaluated demographics, procedural data, technical and procedural success, patency rates and complications. Results: There were 8 (47%) males and the patients' mean age was 66 (± 5.7) years. The mean age of AVF/AVG was 1605 (± 1099) days. All procedures were performed under local anaesthetic. PMT was performed within a mean time of 40 (±34.3) hours from the presentation. Technical, clinical and procedural success was 15/17 (88%). The thrombolysis agents used were tissue plasminogen activator (52.9%) and urokinase (41.2%). Mean primary patency time was 114 (± 116) days, with a 65% 1-month and 47% 3-month primary patency rates. The mean secondary patency time was 155 (±132) days, with 76% one-month and 65% three-month secondary patency rates, respectively. AVF rupture occurred in 3/17 (18%) cases but did not involve loss of the access circuit. Conclusion: The Cleaner XT™ device is a safe, minimally invasive endovascular tool for PMT in thrombosed AVF/AVG, with relatively high success and low complication rates.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。