Trace-Level Automated Mercury Speciation Analysis

痕量级自动化汞形态分析

阅读:1

Abstract

An automated system for methyl Hg analysis by purge and trap gas chromatography (GC) was evaluated, with comparison of several different instrument configurations including chromatography columns (packed column or capillary), detector (atomic fluorescence, AFS, or inductively coupled plasma mass spectrometry, ICP-MS, using quadrupole and sector field ICP- MS instruments). Method detection limits (MDL) of 0.042 pg and 0.030 pg for CH(3)Hg(+) were achieved with the automated Hg analysis system configured with AFS and ICPMS detection, respectively. Capillary GC with temperature programming was effective in improving resolution and decreasing retention times of heavier Hg species (in this case C(3)H(7)Hg(+)) although carryover between samples was increased. With capillary GC, the MDL for CH(3)Hg(+) was 0.25 pg for AFS detection and 0.060 pg for ICP-MS detection. The automated system was demonstrated to have high throughput (72 samples analyzed in 8 hours) requiring considerably less analyst time than the manual method for methyl mercury analysis described in EPA 1630.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。