Design methodology for a confocal imaging system using an objective microlens array with an increased working distance

采用具有增大工作距离的物镜微透镜阵列的共聚焦成像系统的设计方法

阅读:1

Abstract

In this study, a design methodology for a multi-optical probe confocal imaging system was developed. To develop an imaging system that has the required resolving power and imaging area, this study focused on a design methodology to create a scalable and easy-to-implement confocal imaging system. This system overcomes the limitations of the optical complexities of conventional multi-optical probe confocal imaging systems and the short working distance using a micro-objective lens module composed of two microlens arrays and a telecentric relay optical system. The micro-objective lens module was fabricated on a glass substrate using backside alignment photolithography and thermal reflow processes. To test the feasibility of the developed methodology, an optical system with a resolution of 1 μm/pixel using multi-optical probes with an array size of 10 × 10 was designed and constructed. The developed system provides a 1 mm × 1 mm field of view and a sample scanning range of 100 μm. The optical resolution was evaluated by conducting sample tests using a knife-edge detecting method. The measured lateral resolution of the system was 0.98 μm.