Design, Fabrication, and Characterization of Capacitive Micromachined Ultrasonic Transducers for Transcranial, Multifocus Neurostimulation

用于经颅多焦点神经刺激的电容式微机械超声换能器的设计、制造和表征

阅读:1

Abstract

In a recent study using 3-D fullwave simulations, it was shown for a nonhuman primate model that a helmet-shaped 3D array of 128 transducer elements can be assembled for neurostimulation in an optimized configuration with the accommodation of an imaging aperture. Considering all acoustic losses, according to this study, for a nonhuman primate skull, the assembly of the proposed transducers was projected to produce sufficient focusing gain in two different focal positions at deep and shallow brain regions, thus providing sufficient acoustic intensity at these distinct focal points for neural stimulation. This array also has the ability to focus on multiple additional brain regions. In the work presented here, we designed and fabricated a single 15 mm diameter capacitive micromachined ultrasonic transducer (CMUT) element operating at 800 kHz central frequency with a 480 kHz 3 dB bandwidth, capable of producing a 190 kPa peak negative pressure (PNP) on the surface. The corresponding projected transcranial spatial peak pulse average intensity (I(SPPA)) was 28 Wcm(-2), and the mechanical index (MI) value was 1.1 for an array of 128 of these elements.