Validation of a Software Upgrade in a Monte Carlo Treatment Planning System by Comparison of Plans in Different Versions

通过比较不同版本中的计划来验证蒙特卡罗治疗计划系统中的软件升级

阅读:1

Abstract

PURPOSE: Validation of a new software version of a Monte Carlo treatment planning system through comparing plans generated by two software versions in volumetric-modulated arc therapy (VMAT) for lung cancer. MATERIALS AND METHODS: Three patients who were treated with 60 Gy/30 fractions in Elekta Synergy™ linear accelerator by VMAT technique with 2% statistical uncertainty (SU) were chosen for the study. Multiple VMAT plans were generated using two different software versions of Monaco treatment planning system TPS (V5.10.02 and V5.11). By keeping all other parameters constant, originally accepted plans were recalculated for the SUs of 0.5%, 1%, 2%, 3%, 4%, and 5%. For plan evaluation, the metrics compared were conformity Index (CI), homogeneity Index (HI), dose coverage to planning target volume (PTV), organ at risk (OAR) doses to spinal cord, pericardium, bilateral lungs-PTV, esophagus, liver, normal tissue integral dose (NTID), volumes receiving dose >5 and >10 Gy, calculation time (tCT), and gamma pass rates. RESULTS: In both versions, CI and HI improved as the SU increased from 0.5% to 5%. No significant dose difference was observed in Dmean to PTV, bilateral lungs-PTV, pericardium, esophagus, liver, normal tissue volume receiving >5, and >10 Gy and NTID. It was observed that while the tCT and gamma pass rates decreased, the maximum dose to PTV increased as the SU increased. No other significant dose differences were observed between the two MC versions compared. CONCLUSION: For lung VMAT plans, in both versions, SU could be accepted up to 3% per plan with reduced tCT without compromising plan quality and deliverability by accepting variations in point dose and an inhomogeneous dose within the target. The plan quality of Monaco™V5.10.02 was similar to Monaco™TPS-V5.11 except for tCT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。