Abstract
Cerenkov luminescence endoscopy (CLE) is an optical technique that captures the Cerenkov photons emitted from highly energetic moving charged particles (β(+) or β(-)) and can be used to monitor the distribution of many clinically available radioactive probes. A main limitation of CLE is its limited sensitivity to small concentrations of radiotracer, especially when used with a light guide. We investigated the improvement in the sensitivity of CLE brought about by using a β(-) radiotracer that improved Cerenkov signal due to both higher β-particle energy and lower γ noise in the imaging optics because of the lack of positron annihilation. METHODS: The signal-to-noise ratio (SNR) of (90)Y was compared with that of (18)F in both phantoms and small-animal tumor models. Sensitivity and noise characteristics were demonstrated using vials of activity both at the surface and beneath 1 cm of tissue. Rodent U87MG glioma xenograft models were imaged with radiotracers bound to arginine-glycine-aspartate (RGD) peptides to determine the SNR. RESULTS: γ noise from (18)F was demonstrated by both an observed blurring across the field of view and a more pronounced fall-off with distance. A decreased γ background and increased energy of the β particles resulted in a 207-fold improvement in the sensitivity of (90)Y compared with (18)F in phantoms. (90)Y-bound RGD peptide produced a higher tumor-to-background SNR than (18)F in a mouse model. CONCLUSION: The use of (90)Y for Cerenkov endoscopic imaging enabled superior results compared with an (18)F radiotracer.