Simulated low-dose multi-detector computed tomography: spatial effects on surrogate parameters of bone strength at the proximal femur

模拟低剂量多层螺旋CT:空间效应对股骨近端骨强度替代参数的影响

阅读:1

Abstract

This study investigated simulated tube current reduction and sparse sampling for low-dose computed tomography (CT) regarding volumetric bone mineral density (vBMD) and cortical bone thickness (Ct.Th) of the proximal femur. Sparse sampling with dose reductions of up to 90% may still allow extraction of bone strength parameters with clinically acceptable accuracy. INTRODUCTION: We aimed to investigate effects of CT with simulated lowered tube current and sparse sampling on trabecular and cortical vBMD as well as Ct.Th of the entire proximal femur, its subregions, and with detailed spatial assessments. METHODS: Clinical routine multi-detector CT (MDCT) scans covering the hips from 40 patients were used for simulations of low-dose imaging with 50% and 10% of the original tube current (D50, D10) or projections (P50, P10) combined with statistical iterative reconstruction (SIR), which were then compared against original data with full dose (D100 P100) regarding trabecular vBMD, cortical vBMD, and Ct.Th. An automated framework for multi-parametric assessments was used. Relative errors by comparing measures from original data and simulated low-dose data, regression analyses, Bland-Altman analyses, and statistical parametric mapping (SPM, to assess the spatial distribution of accuracy) were computed. RESULTS: Sparse sampling enabled drastic reductions of radiation exposure (down to 10% of original imaging) while still producing determinants of bone strength with clinically acceptable relative changes. Lower biases according to Bland-Altman analyses were observed for sparse sampling compared to imaging with virtually lowered tube currents (D10 P100 versus D100 P10) regarding trabecular vBMD, cortical vBMD, as well as Ct.Th. Better accuracy across the whole proximal femur for D100 P50 than for D50 P100 and for D100 P10 than for D10 P100 was observed. CONCLUSIONS: Sparse sampling with SIR may enable drastic reductions of radiation exposure (up to 90% of original doses) for opportunistically measuring image-based surrogate parameters of bone strength.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。