Safety and efficacy of Surpass Evolve Flow diverter for intracranial aneurysms: A study of 116 patients

Surpass Evolve血流导向装置治疗颅内动脉瘤的安全性和有效性:一项纳入116例患者的研究

阅读:2

Abstract

BACKGROUND AND PURPOSE: Flow diverter embolization is a recognized method for treating intracranial aneurysms. This study evaluates the safety and efficacy of the Surpass Evolve flow diverter in treating intracranial aneurysm. MATERIAL AND METHODS: From May 2019 to June 2022, our center prospectively enrolled patients presenting with both ruptured and unruptured intracranial aneurysms. We assessed aneurysm occlusion, along with the occurrence of ischemic and hemorrhagic complications, and mortality at 6-months. The 3-month occlusion rate was determined using MR angiography, while the 6-month occlusion rate employed the O'Kelly-Marotta (OKM) grading scale on DSA. RESULTS: A total of 116 patients with 120 aneurysms received treatment with the Surpass Evolve device. The average aneurysm size was 6.6 mm (range: 2-30 mm). All patients were administered two loading doses of ticagrelor (180 mg) one day before the procedure and 2 h pre-embolization. A procedural complication occurred in one case, involving wire-related perforation of an M3 branch distant from the giant aneurysm; however, this evolved favorably. There were no reported deaths linked to the treatment. Permanent neurological deficits were observed in 3 (2.5%) patients, primarily due to early in-stent thrombosis. MR angiography results at 3 months indicated complete occlusion for 87 out of 115 (75.6%) monitored aneurysms. The 6-month DSA showed neck remnant or complete occlusion (OKM grade C or D) in 86 out of the 107 (80.4%) patients who underwent follow-up. CONCLUSION: The Surpass Evolve stands out as a reliable and effective flow diverter for the management of intracranial aneurysms.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。