BACKGROUND: Endovascular embolization of wide-necked aneurysms can be challenging. The development of intrasaccular devices like the Contour has enabled us to approach these aneurysms effectively by reducing recanalization rates and eliminating the need for dual antiplatelet therapy, which is particularly beneficial in the case of ruptured aneurysms. Although complications from using these devices are rare, it is crucial to address them properly. In this case, the authors highlight how to manage artery thrombosis caused by device protrusion during aneurysm embolization. OBSERVATIONS: This report describes a complication in a male patient with a ruptured anterior communicating artery wide-necked aneurysm. Following Contour-assisted coiling of the aneurysm, a realignment of the detachable apex of the device occluded the A2 segment of the right anterior cerebral artery. After the failure of intra-arterial and intravenous tirofiban infusion as well as mechanical thrombectomy, a self-expanding open-cell stent was deployed in the involved vessel, achieving successful reperfusion. LESSONS: The Contour device has a detachable zone that can cause occlusion of the parent vessel after deployment. The use of a stent as a rescue maneuver may be useful if reperfusion of the vessel cannot be achieved through other methods such as aspiration or full-dose antiplatelet therapy.