Removal of Antiscatter Grids for Spinal Digital Subtraction Angiography: Dose Reduction without Loss of Diagnostic Value

脊柱数字减影血管造影中去除防散射栅格:降低辐射剂量而不损失诊断价值

阅读:1

Abstract

Background Spinal digital subtraction angiography (DSA) exposes patients and operators to substantial amounts of radiation. Antiscatter grid (ASG) removal is used to decrease radiation exposure but may reduce image quality. Purpose To determine whether ASG removal during spinal DSA in adults reduces radiation dose while maintaining diagnostic image quality and whether dose reduction is related to body mass index (BMI). Materials and Methods This Health Insurance Portability and Accountability Act-compliant prospective study included adults undergoing spinal DSA between January and December 2016. Each procedure included an additional angiographic acquisition performed twice, once with and once without ASG, either documenting the artery of Adamkiewicz (no pathology group) or the condition leading to the procedure (pathology group). Dose differences between study acquisitions and the influence of BMI were evaluated via paired t test. Two neurointerventionalists blinded to acquisition protocols were asked to independently evaluate a sample of 40 study acquisitions (20 with ASG, 20 without ASG) from 20 randomly selected participants to (a) rate image quality, (b) categorize findings, and (c) determine whether images had been obtained with or without ASG. Percentage agreement on image quality, findings categorization, and ability to correctly identify the acquisition protocol was calculated for both readers. Results Fifty-three participants (mean age ± standard deviation, 51 years ± 15.2; 32 men) were evaluated. ASG removal reduced the mean dose per acquisition by approximately 33% (mean dose-area product and air kerma decreased from 202 to 135.6 µGy/m(2) and from 35.3 to 24 mGy, respectively; P < .001) independently of BMI (P = .3). Both readers evaluated all images (40 of 40) as being of diagnostic quality and correctly categorized findings in 19 of 20 (95%) cases. Overall percentage agreement for correct protocol identification was 60% (12 of 20) for grid-in and 45% (nine of 20) for grid-out images. Conclusion Antiscatter grid removal during spinal digital subtraction angiography decreased participants' radiation exposure while preserving diagnostic image quality. © RSNA, 2020.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。