Evaluating the Tubridge™ flow diverter for large cavernous carotid artery aneurysms

评估 Tubridge™ 血流导向装置在治疗大型海绵窦段颈内动脉瘤中的应用

阅读:1

Abstract

BACKGROUND: The Tubridge™ flow diverter (TFD) was recently developed in China; however, its safety and efficacy in treating large cavernous carotid artery aneurysms (LCCAs) are unclear. Our objective was to evaluate the safety and efficacy of the TFD in patients receiving TFDs to treat LCCAs (10-25 mm). METHODS: Between June 2013 and May 2014, seven patients with LCCAs were enrolled in our study, and all seven patients underwent TFD implantation combined with coils. RESULTS: Angiographic follow-up images were available for all seven patients at a median of 57.5 ± 16.7 (range, 6-69) months. Seven patients obtained favorable angiographic results defined as O'Kelly-Marotta Scale C and D. Clinical follow-up data were available for all seven patients at a median of 73.32 ± 3.6 (range, 66-78) months. No patients developed new neurological deficits. Six patients achieved a modified Rankin scale score of 0, and diplopia improved in the remaining patient. CONCLUSIONS: The results were excellent for the aneurysms treated with TFDs in our patients with LCCAs. TFDs are feasible for the treatment of LCCAs, but a multicenter, controlled clinical trial is needed to evaluate the long-term safety and efficacy of the TFD to treat LCCAs.