Accidental guide catheter fracture in mechanical thrombectomy

机械取栓术中导引导管意外断裂

阅读:1

Abstract

Endovascular thrombectomy is now the standard of care for large vessel occlusion stroke. The aim is to achieve rapid and complete recanalisation while avoiding complications. Apart from the conventional complications of neurointerventional procedures, mechanical thrombectomy has its unique set of complications, inherent to the disease pathophysiology. We describe an unusual complication of catheter fracture and subsequent distal embolisation into the cerebral vasculature, which was noticed 24 hours after the procedure. Due to a lack of clinical consequences, we decided to manage it conservatively. The patient died within the following few days from respiratory complications unrelated to the stroke or the endovascular thrombectomy procedure. Consequently, we were able to retrieve the fractured segment and carry out histopathological analysis, which helped us to identify exactly its origin from the guide catheter. We believe that systematic reporting and database compilation of such device-related complications will aid in the design and development of neurointerventional devices in the future.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。