We evaluated Computed Tomography Angiography (CTA) of four different self-expandable stents in phantom model. Three stents ware made of nitinol and one was made of cobaltbased alloy. Each stent placed in plastic injector, filled with diluted iodinated contrast material, and placed on a plastic case filled with saline. CTA was performed on a four-detector CT scanner (detector collimation, 0.5mm: helical pitch 3.5). Axial and longitudinal reconstructions of CT imaging were compared with the instent attenuation. Three nitinol stents showed few artifact, and cobalt-based alloy stent was appeared artificial narrowing. In conclusion, CTA is useful for follow-up examination after carotid artery stenting (CAS) as a less invasive evaluation method.