Pilot Application of Magnetic Nanoparticle-Based Biosensor for Necrotizing Enterocolitis

基于磁性纳米粒子的生物传感器在坏死性小肠结肠炎中的初步应用

阅读:8
作者:Dokyoon Kim #, Changlin Fu #, Xuefeng B Ling #, Zhongkai Hu, Guozhong Tao, Yingzhen Zhao, Zachary J Kastenberg, Karl G Sylvester, Shan X Wang

Background

Necrotizing Enterocolitis (NEC) is a major source of neonatal morbidity and mortality. There is an ongoing need for a sensitive diagnostic instrument to discriminate NEC from neonatal sepsis. We hypothesized that magnetic nanopartile-based biosensor analysis of gut injury-associated biomarkers would provide such an instrument. Study design: We designed a magnetic multiplexed biosensor platform, allowing the parallel plasma analysis of C-reactive protein (CRP), matrix metalloproteinase-7 (MMp7), and epithelial cell adhesion molecule (EpCAM). Neonatal subjects with sepsis (n=5) or NEC (n=10) were compared to control (n=5) subjects to perform a proof of concept pilot study for the diagnosis of NEC using our ultra-sensitive biosensor platform.

Conclusion

The exploration of the multiplexed nano-biosensor platform shows promise to deliver an ultrasensitive instrument for the diagnosis of NEC in the clinical setting.

Results

Our multiplexed NEC magnetic nanoparticle-based biosensor platform was robust, ultrasensitive (Limit of detection LOD: CRP 0.6 pg/ml; MMp7 20 pg/ml; and EpCAM 20 pg/ml), and displayed no cross-reactivity among analyte reporting regents. To gauge the diagnostic performance, bootstrapping procedure (500 runs) was applied: MMp7 and EpCAM collectively differentiated infants with NEC from control infants with ROC AUC of 0.96, and infants with NEC from those with sepsis with ROC AUC of 1.00. The 3-marker panel comprising of EpCAM, MMp7 and CRP had a corresponding ROC AUC of 0.956 and 0.975, respectively.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。