A comparison of 2 micron inner diameter open tubular column liquid chromatography with pressure-driven isocratic, slip-flow, and electrochromatographic modes of operation: a theoretical study

2微米内径开口管柱液相色谱法在压力驱动等度洗脱、滑移流洗脱和电色谱模式下的比较:一项理论研究

阅读:1

Abstract

Modifications to the flow profile used in open tube capillary liquid chromatography (OT-CLC) include using slip-flow walls and using electroosmosis as a fluid pump as practiced in electrochromatography. These modifications are implemented experimentally by changing the capillary surface and solvent conditions which results in the change of boundary conditions at the capillary wall. In this paper we employ a theory-based study and compare the zone broadening of simple solutes using parabolic flow from a liquid pump, slip-flow from a highly hydrophobic inner surface with water eluent, and electroosmosis for the conditions of pure water and dilute salt utilizing 2 µm inner diameter OT capillaries. In general, the two types of flow other than parabolic exhibit thin zones in the early part of the chromatogram, consistent with previous studies of slip-flow and electroosmotic flow used in electrochromatography. Electrochromatography is shown to yield higher efficiency and less zone broadening than parabolic and slip-flow conditions used in this study. Nonetheless, it is found that the zone standard deviations are shown to be similar for these flow profiles as is the number of plates for these different flow profiles under the conditions utilized here. It is revealed that these modifications do not warrant the effort to maintain the special solvent conditions when compared to gradient elution OT-CLC, which gives a nearly constant peak width throughout the chromatogram, is easiest to implement, and is the method of choice for complex analysis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。