Contact-free scanning and imaging with the scanning ion conductance microscope

利用扫描离子电导显微镜进行非接触式扫描和成像

阅读:1

Abstract

Scanning ion conductance microscopy (SICM) offers the ability to obtain very high-resolution topographical images of living cells. One of the great advantages of SICM lies in its ability to perform contact-free scanning. However, it is not yet clear when the requirements for this scan mode are met. We have used finite element modeling (FEM) to examine the conditions for contact-free scanning. Our findings provide a framework for understanding the contact-free mode of SICM and also extend previous findings with regard to SICM resolution. Finally, we demonstrate the importance of our findings for accurate biological imaging.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。