Unbiased Determination of Adsorption Isotherms by Inverse Method in Liquid Chromatography

利用液相色谱反相法无偏测定吸附等温线

阅读:1

Abstract

The Inverse Method is a widely used technique for the determination of adsorption isotherms in liquid chromatography. In this method, isotherm is determined from the overloaded peak profile of the component by the iterative solution of the mass balance equation of liquid chromatography. Successful use of this method requires a prior assumption of equation of isotherm (Langmuir, BET etc.). In this work, we have developed an inverse method that gives results of similar accuracy to the frontal analysis without assuming the equation of the isotherm. The oversaturated peaks were calculated using a spline fitted to data points instead of the derivative of the isotherm. The distribution of the isotherm points were optimized for minimizing the difference between the measured and calculated overloaded peaks. The accuracy of the developed method was verified with synthetic benchmark peaks and by the determination of isotherm of buthyl-benzoate under real conditions. The results confirmed that the accuracy of the developed method is similar to that of Frontal Analysis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。